Patient-specific biokinetics and hybrid 2D/3D approach integration in OEDIPE software: application to radioiodine therapy - UNICANCER Accéder directement au contenu
Article Dans Une Revue Physica Medica Année : 2023

Patient-specific biokinetics and hybrid 2D/3D approach integration in OEDIPE software: application to radioiodine therapy

Résumé

Background The development of targeted radionuclide therapy requires the development of dosimetry software accounting for patient-specific biokinetics. New functionalities were thus developed in the OEDIPE software, to deal with multiple 3D images or multiple planar images and a SPECT image. Materiel & Method Methods were implemented to recover patient biokinetics in volumes of interest. If several 3D SPECT images are available, they are registered to a reference CT scan. When several planar images and a single SPECT are available, the planar images are registered to the SPECT and counts of the planar images converted to activity. To validate these developments, six SPECT/CT and planar images of a Jaszczak phantom containing I-131 were acquired at different dates. Cumulated activity was estimated in each sphere using the SPECT/CT images only or the planar series associated to one SPECT/CT. Biokinetic and doses in lesions and in the lungs of a patient treated with I-131 for differentiated thyroid cancer were then estimated using four planar images and a SPECT/CT scan. Whole-body retention data were used to compare the biokinetics obtained from the planar and SPECT data. Results Activities and cumulated activities estimated using OEDIPE in the phantom spheres agreed well with the reference values for both approaches. Results obtained for the patient compared well with those derived from whole-body retention data. Conclusion The implemented features allow automatic evaluation of patient-specific biokinetics from different series of patient images, enabling patient-specific dosimetry without the need for external software to estimate the cumulated activities in different VOIs.
Fichier principal
Vignette du fichier
Bensiali_EJMP22.pdf (593.88 Ko) Télécharger le fichier
Figure4_Net_UpdateWithR2.pdf (89.15 Ko) Télécharger le fichier
Figure_1.pdf (219.21 Ko) Télécharger le fichier
Figure_2.pdf (89.53 Ko) Télécharger le fichier
Figure_3.pdf (60.19 Ko) Télécharger le fichier
Figure_5.pdf (101.49 Ko) Télécharger le fichier
Figure_6.pdf (117.52 Ko) Télécharger le fichier
Figure_7.pdf (88.85 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
licence : CC BY NC ND - Paternité - Pas d'utilisation commerciale - Pas de modification
licence : Copyright (Tous droits réservés)
Origine : Fichiers produits par l'(les) auteur(s)
licence : Copyright (Tous droits réservés)
licence : Copyright (Tous droits réservés)
licence : Copyright (Tous droits réservés)
licence : Copyright (Tous droits réservés)
licence : Copyright (Tous droits réservés)
licence : Copyright (Tous droits réservés)

Dates et versions

hal-03960724 , version 1 (05-06-2023)

Licence

Paternité - Pas d'utilisation commerciale - Pas de modification

Identifiants

Citer

Mohammed Bensiali, Nadège Anizan, Sophie Leboulleux, Stephanie Lamart, Estelle Davesne, et al.. Patient-specific biokinetics and hybrid 2D/3D approach integration in OEDIPE software: application to radioiodine therapy. Physica Medica, 2023, 113, pp.102462. ⟨10.1016/j.ejmp.2022.09.013⟩. ⟨hal-03960724⟩
105 Consultations
71 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More